FTR-1 Foot Tester
Features
- Test range of 0.9 ~ 35 mega Ohms
- Dual footwear station for repeatable checks on both right and left
feet at the same time
- Audible and visual indication of system status
- Light indicator ( red, green, yellow)
- Sturdy plastic coated metal frame with wall mounting bracket as standard
Calibration
Calibration is recommended on average of every twelve months. A resistance
decade box can be used. Connect on of the two left stainless steel
foot print. Set the resistance just below the lowest acceptable resistance
level of the tester. The RED low fail LED will light and an audible
buzzer will pulse on the left side of the tester, when you push the "Press
to Test" button. Increase the resistance of your right foot,
connect the decade box lead currently making contact with the left
stainless
steel foot print, over to the right foor print and repeat the process.
Part
Number : SCJ/EQ-FTR-1
| Power
Supply |
9V
battery |
Test
Voltage |
Nominal
9V |
Temperature
Range
Operating
Storage |
5-59°C
(-15 to 60°C) |
Relative
Humidity |
0%
- 90% (non-condensing) |
Accuracy |
+/-
5% |
Repeatability
|
+/-
5% |
Weight |
4.5
kg |
Dimensions |
190
x 135 x 45mm |
Height
of station |
1m
- 39 1/2" |
Light
Indicator |
HIGH
FAIL
PASS
LOW FAIL
LOW BATTERY |
RED
- System resistance in excess 35 megaOhm
GREEN - acceptable resistance
RED- System resistance less than 0.9 megaOhms
YELLOW |